Model Number UHI-4
Device Problem Restricted Flow rate (1248)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported to olympus that the high flow insufflation unit was in use and did not supply gas at desired rate (45 liters/minute).The customer reported that the delivery rate was measured at 0.3 liters/minute, 0.6 liters/minute, 6 liters per minute and 11 liters per minute.The customer stated the delivery rate was not a problem for their other insufflator devices.The customer reported the device was supplied by the hospital medical wall gas system.No adverse effects to patient reported.
 
Manufacturer Narrative
Additional information has been requested regarding this event.No response has been received at this time.The device has not been returned for evaluation.A supplemental report will be filed when the investigation is completed.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device was returned to olympus for inspection, and the customer's complaint was not confirmed.During the inspection, there were no faults/findings.The device meets all inspection criteria.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause was unable to be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17777456
MDR Text Key323811341
Report Number3002808148-2023-10065
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/22/2023
Initial Date FDA Received09/19/2023
Supplement Dates Manufacturer Received11/20/2023
Supplement Dates FDA Received11/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-