EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PHYSIO II ANNULOPLASTY RING; RING, ANNULOPLASTY Back to Search Results
Model Number 5200M
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Endocarditis (1834)
Event Date 04/26/2023
Event Type  Injury  
Event Description
Edwards lifesciences maintains an implant patient registry (ipr).This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional "customer complaint".The information reported may or may not be related to the edwards device.Edwards received notification via the implant patient registry that this ring model 5200m26 was explanted from the mitral position after an implant duration of one (1) month and five (5) days due to mitral valve infective endocarditis.A 27mm magna ease mitral valve was implanted in replacement.Unfortunately, patient passed away one (1) month and thirteen (13) days after the reintervention.
 
Manufacturer Narrative
H10: additional manufacturer narrative: the subject device will not be returned for evaluation due to patient's infective endocarditis.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Manufacturer Narrative
H10: additional manufacturer narrative: the device history record (dhr) was reviewed and shows that this device met all manufacturing specifications for product release prior to distribution.No issues were identified that would have impacted this event.Despite due diligent attempts to receive further information regarding this event, no additional information was received from the customer.Prosthetic endocarditis, with or without vegetation, of valves and annuloplasty rings is a serious complication of cardiac valve replacement and valve repair surgeries despite improvements in prostheses types, surgical techniques, and infection control measures.Prosthetic endocarditis occurring within 60 days of valve or ring implantation generally reflects contamination arising in the perioperative period.There are many opportunities for organisms to seed a prosthesis perioperatively, most of which probably occurs intraoperatively.Besides the patients own skin and access lines, several other important modes of contamination have been recognized including air in the operating room, the coronary suction devices used during surgery and the heart-lung bypass machine, faulty technique during cardiac output measurements, and the prosthesis itself.In early cases of prosthetic endocarditis, subsequent infections are almost universally related to contamination at the time of surgery.If there were ever non-conformances in the sterility or packaging processes, they would most likely manifest in the early post-operative period.Since there are multiple modes of contamination other than the prosthesis itself, and no indication or allegation that the patients infection was device related, the event was likely due to patient and/or procedural-related factors.A definitive root cause cannot be conclusively determined; however, patient and/or procedural factors likely caused or contributed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PHYSIO II ANNULOPLASTY RING
Type of Device
RING, ANNULOPLASTY
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
one edwards way
mle fl2 office m2013
irvine, CA 92614
9492506615
MDR Report Key17779465
MDR Text Key323786312
Report Number2015691-2023-16050
Device Sequence Number1
Product Code KRH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K083470
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number5200M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/29/2023
Initial Date FDA Received09/20/2023
Supplement Dates Manufacturer Received10/23/2023
Supplement Dates FDA Received10/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Hospitalization; Required Intervention;
Patient Age77 YR
Patient SexFemale
-
-