• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION UNI-VENT, MODEL 731; VENTILATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION UNI-VENT, MODEL 731; VENTILATOR Back to Search Results
Model Number 799-EGB2-00-01
Device Problem Communication or Transmission Problem (2896)
Patient Problem Cardiac Arrest (1762)
Event Date 07/24/2023
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event Description
Complainant alleged that while attempting to treat a 60-year-old male patient, the device displayed "internal comm failure - 1173," "internal comm failure-1471" and "internal comm failure - 1475" error messages.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing without duplicating the report.The input adaptor was replaced as a pre-caution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNI-VENT, MODEL 731
Type of Device
VENTILATOR
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key17780002
MDR Text Key323797951
Report Number1220908-2023-03545
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00847946024208
UDI-Public00847946024208
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111473
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number799-EGB2-00-01
Device Catalogue Number799-EGB2-00-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/30/2023
Initial Date FDA Received09/20/2023
Supplement Dates Manufacturer Received08/30/2023
Supplement Dates FDA Received10/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age60 YR
Patient SexMale
-
-