Model Number 799-EGB2-00-01 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
Cardiac Arrest (1762)
|
Event Date 07/24/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that while attempting to treat a 60-year-old male patient, the device displayed "internal comm failure - 1173," "internal comm failure-1471" and "internal comm failure - 1475" error messages.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing without duplicating the report.The input adaptor was replaced as a pre-caution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
|