MAQUET CARDIOVASCULAR LLC HEMOPRO POWER SUPPLY; UNIT, CAUTERY, THERMAL, AC-POWERED Back to Search Results
Model Number C-VH-3010
Device Problem Failure to Deliver Energy (1211)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/05/2023
Event Type  malfunction  
Event Description
The hospital reported that during preparation for an endoscopic vein harvesting procedure, t.W.Power supply s/n (b)(6) was an out of box failure.Power supply was cleared through biomed but then failed to provide power once placed in the or.Light flickers when turned on and multiple cables were tried to get power to evh with no success.Device was not in contact with patient.
 
Manufacturer Narrative
Tw id# 882412.The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Event Description
N/a.
 
Manufacturer Narrative
Trackwise # (b)(4).Updated section: b4, d10, g4, g7, h2, h3, h6, h10.The device was returned to the factory for evaluation on 09/18/2023.An investigation was conducted on 09/06/2023.A visual inspection was conducted.Signs of clinical use and no evidence of blood were observed on the device.The power supply was observed to be intact.An electrical evaluation was conducted.A reference power cord was plugged into the power supply and the device was switched on.The green lights on the power supply were faintly illuminated, indicating insufficient power was delivered to the device.A pre-cautery test was performed per the instruction for use (ifu) with a reference adapter, hp2, cable, and power supply vh-3010 at level 3.0.The device did not pass the pre-cautery test.No visible steam or heat was produced when the device was activated and no tone was audible from the power supply upon activation.Final testing of the device was conducted on 09/26/2023 using a a fluke 8846a precision multimeter bmram id# (b)(6), per mcv00030545 rev.B, equipment calibration procedure for vasoview power supply.The following results were observed: no load voltage: 2.58vdc; low current output: 0.0 amps dc; high current output: 0.0 amps dc.The voltage and current were found to be out of specification.Based on the returned condition of the device, the reported failure "failure to deliver energy" was confirmed.The certificate of conformance (c of c) was reviewed.The vendor certifies that this device serial conforms to all applicable product specifications.The reported device is an oem device.The certificate of conformance was reviewed for the serial # (b)(6).The vendor certifies that this device serial # conforms to all applicable product specifications and there were no non-conformances identified for the manufacturing batch.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOPRO POWER SUPPLY
Type of Device
UNIT, CAUTERY, THERMAL, AC-POWERED
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key17780309
MDR Text Key323824856
Report Number2242352-2023-00794
Device Sequence Number1
Product Code HQO
Combination Product (y/n)N
PMA/PMN Number
K052274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 10/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC-VH-3010
Device Catalogue NumberC-VH-3010
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 08/29/2023
Initial Date FDA Received09/20/2023
Supplement Dates Manufacturer Received10/30/2023
Supplement Dates FDA Received10/30/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2023
Is the Device Single Use? No
Type of Device Usage Initial
Patient Sequence Number1
-
-