REMOTE DIAGNOSTIC TECHNOLOGIES LTD. TEMPUS PRO; MONITOR, PHYSIOLOGICAL, PATIENT (WITH ARRHYTHMIA DETECTION OR ALARMS)
|
Back to Search Results |
|
Model Number 00-1026-R |
Device Problem
Operating System Becomes Nonfunctional (2996)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It has been reported that the device is having software issues.
|
|
Manufacturer Narrative
|
During further investigation by the bench technician, it was observed that the tempus pro intermittently displays a software error after rebooting.The bench technician stated the sd card was corrupt and replaced it address the issue.The device was returned to full functionality.No further evaluation is warranted at this time.Based on the information available and the testing conducted, the cause of the reported problem was a corrupt sd card.The reported problem was confirmed.Based on the information available and results of additional analysis, no further action is necessary at this time.The device was operational after repairs were completed.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|
|
Manufacturer Narrative
|
Evaluation method code grid, evaluation results code grid, component code grid to reflect pr (b)(4).
|
|