The hillrom technician found the cpr valve needed to be replaced.Per the hillrom service manual the totalcare bariatric bed and totalcare bed system requires an effective maintenance program.We recommend that you perform a semi-annual preventative maintenance.Test the cpr release for correct operation and reset of the head cylinder.A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed on (b)(6), 2023.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the cpr valve to resolve the reported event.Based on this information, no further action is required.
|