ORTHO-CLINICAL DIAGNOSTICS, INC. VITROS CHEMISTRY PRODUCTS GLU SLIDES; IN VITRO DIAGNOSTICS Back to Search Results
Model Number 1707801
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/23/2023
Event Type  malfunction  
Event Description
The customer reported non-reproducible, higher than expected vitros glu results were obtained from two different patient samples tested on a vitros xt7600 integrated system.Patient 1: vitros glu results of >625, >1250, 402 mg/dl vs.The expected result of 178.0 mg/dl.Patient 2: vitros glu result of >625, >1250 mg/dl vs.The expected result of 68.0 mg/dl.Biased results of the direction and magnitude observed could lead to inappropriate physician action if they were to occur undetected on patient samples.The higher than expected vitros glu patient results were not reported from the laboratory.There were no allegations of patient harm as a result of this event.This report is number two of five mdr¿s for this event.Five 3500a forms are being submitted for this event as five devices were involved.This report corresponds to ortho clinical diagnostics inc (ortho) complaint number (b)(4) and reportability assessment (b)(4).
 
Manufacturer Narrative
The investigation determined that non-reproducible, higher than expected vitros glu results were obtained from two different patient samples tested on a vitros xt7600 integrated system.The assignable cause for this event could not be determined.The higher than expected vitros glu results for both patient samples and the expected glu results were obtained from the same vitros glu slide cartridge.It is possible that there is a potential performance issue with some, but not all, vitros glu slides from the affected slide cartridge.Based on a review of historical quality control results, there was no indication of a performance issue at the customer site with vitros glu slide lot 0023-3233-9086.However, continual track and trending of complaints identified a slight increase in complaints for higher than expected and non-reproducible results using vitros glu slide lot 0023-3233-9086, and with vitros glu slides from coating 3233.Therefore, a vitros glu slide lot issue could not be entirely ruled out as a contributor to the event.No diagnostic within run precision testing was performed to assess the performance of the vitros xt7600 integrated system.Therefore, it is unknown if the vitros xt7600 integrated system was performing as intended and an instrument issue cannot be ruled out as a contributor to the event.Pre-analytical sample processing could not be ruled out as a contributing factor, as it is unknown if the customer was following the sample collection device manufacture¿s recommendation for sample centrifugation and cellular debris, due to poor sample preparation, could have been present in the affected sample, although this could not be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITROS CHEMISTRY PRODUCTS GLU SLIDES
Type of Device
IN VITRO DIAGNOSTICS
Manufacturer (Section D)
ORTHO-CLINICAL DIAGNOSTICS, INC.
100 indigo creek drive
rochester NY 14626
Manufacturer (Section G)
ORTHO-CLINICAL DIAGNOSTICS, INC.
513 technology boulevard
rochester NY 14626
Manufacturer Contact
laurie o'riordan
microtyping systems
1295 southwest 29th avenue
pompano beach, FL 33069
9549709500
MDR Report Key17783391
MDR Text Key323863012
Report Number0001319809-2023-00118
Device Sequence Number1
Product Code CGA
Combination Product (y/n)N
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1707801
Device Catalogue Number1707801
Device Lot Number0023-3233-9086
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/23/2023
Initial Date FDA Received09/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/28/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-