• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S SUPRIS RETROPUBIC KIT; FEMALE STRESS URINARY INCONTINENCE SURGICAL MESH-SLING, SYNTHETIC POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S SUPRIS RETROPUBIC KIT; FEMALE STRESS URINARY INCONTINENCE SURGICAL MESH-SLING, SYNTHETIC POLYMER Back to Search Results
Catalog Number 51956
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/28/2023
Event Type  malfunction  
Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release and no trends were noted.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
 
Event Description
According to available information, this device required revision due to a tear.The doctor tore the device while passing the introducer.The procedure was successfully completed but the device had to be modified.No other adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPRIS RETROPUBIC KIT
Type of Device
FEMALE STRESS URINARY INCONTINENCE SURGICAL MESH-SLING, SYNTHETIC POLYMER
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key17784519
MDR Text Key323899431
Report Number2125050-2023-01154
Device Sequence Number1
Product Code OTN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111233
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 09/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number51956
Device Lot Number9005479_5195622400
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/20/2023
Date Device Manufactured09/24/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-