Catalog Number B12LT
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/01/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date sent: 9/20/2023.B3: unknown, assumed first day of month that complaint was reported.D4: batch # x70315.Investigation summary.The product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that the b12lt device was received with the tip of the sleeve damaged melted.In addition, the packaging opened was returned along with the instrument.The event reported was confirmed and it is related to improper use of the device.In order to prevent this kind of damage please avoid the contact with laser, electrosurgical or ultrasonic devices.As part of ethicon¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
 
Event Description
It was reported that during an unknown procedure, defective tip on edge of trocar sleeve.It was noticed upon entry of trocar into the abdominal cavity.No patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOPATH XCEL BLADELESS TROCAR
Type of Device
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key17784927
MDR Text Key323873793
Report Number3005075853-2023-06724
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K032676
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 09/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberB12LT
Device Lot Number539C31
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2023
Initial Date Manufacturer Received 08/26/2023
Initial Date FDA Received09/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/04/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-