Model Number CYF-5A
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/24/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for a device evaluation and the customer's allegation was confirmed.In addition to the reported in b5, the device evaluation found the following: the bending section cover rubber had cut and slack, due to a pinhole on bending section cover rubber the water tightness was lost, the connecting tube had a scratch and a wrinkle, the eyepiece had a scratch, the grip had a scratch, and the lock engagement lever had a scratch.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if additional information becomes available.
 
Event Description
The customer reported to olympus that the oes cystonephrofiberscope had a leak.The issue was observed during a diagnostic (cystoscopy) procedure.The procedure was completed with a similar device and there were no reports of patient or user harm associated with this event.The device was returned to olympus for a device evaluation and found the bending section cover rubber fell off and was missing.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the falling section cover rubber could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES CYSTONEPHROFIBERSCOPE
Type of Device
CYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17786294
MDR Text Key323902719
Report Number3002808148-2023-10125
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K032092
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 09/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberCYF-5A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/20/2023
Supplement Dates Manufacturer Received09/24/2023
Supplement Dates FDA Received09/29/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/12/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-