• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN PROTEGE EVERFLEX SELF-EXPANDING STENT SYSTEM LONG; STENTS, DRAINS AND DILATORS FOR THE BILIARY DUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN PROTEGE EVERFLEX SELF-EXPANDING STENT SYSTEM LONG; STENTS, DRAINS AND DILATORS FOR THE BILIARY DUCTS Back to Search Results
Model Number PRB35-06-150-120
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Unspecified Infection (1930)
Event Date 08/16/2023
Event Type  Injury  
Event Description
During the index procedure two protege everflex stents were implanted in the right proximal, middle and distal superficial femoral artery (sfa) (r1).Approximately 7 months post procedure the patient suffered from a right thigh incision infection.The patient was seen in the clinic for follow up for a previous bypass surgery.A thigh incision infection was noted and the patient was sent to hospital and underwent a computed tomography angiography (cta).The patient was treated with antibiotics and was taken to the operating room for wound debridement and a wound vac was placed.The patient recovered.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTEGE EVERFLEX SELF-EXPANDING STENT SYSTEM LONG
Type of Device
STENTS, DRAINS AND DILATORS FOR THE BILIARY DUCTS
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key17787014
MDR Text Key323894280
Report Number2183870-2023-00347
Device Sequence Number1
Product Code FGE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberPRB35-06-150-120
Device Catalogue NumberPRB35-06-150-120
Device Lot NumberB468335
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/23/2023
Initial Date FDA Received09/21/2023
Date Device Manufactured10/26/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
PROTEGE EVERFLEX PRB35-06-200-120 LOT#A985587.
Patient Outcome(s) Required Intervention; Hospitalization; Other;
Patient Age72 YR
Patient SexFemale
Patient Weight55 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-