It was reported that the coil was fractured.The target lesion was located in the splenic artery with multiple aortic aneurysms.A 10mm x 40cm.035 interlock cube coil was selected for use.During the procedure, the device was delivered normally into the non-boston scientific contrast catheter.However, when the physician continued to push the coil, the coil got stuck and broke at the opening of the contrast catheter.The coil could not be pulled back.The physician withdraws the coil together with the delivery system.The contrast between the coil and the catheter was found stretched at the place of the fracture.The physician tried to continue to use the device and cut off the stretched and broken part and tried to place the released coil, which was exposed outside of the delivery system, into the contrast catheter, but ultimately failed.The procedure was completed with a different device.No complications were reported, and the patient was stable.
|