Model Number 900PT561
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/21/2023
Event Type  malfunction  
Event Description
A distributor reported on behalf of a healthcare facility in china that the autofill chamber as part of the 900pt561 heated breathing tube and chamber kit was found with a vertical crack on the chamber dome.Upon device set-up, the chamber was observed to be leaking in large quantities.There was no patient involvement as the issue was found whilst the device was not in use on a patient.
 
Manufacturer Narrative
(b)(4).Chamber kit is a component designed for use with the airvo 2 humidification series to provide nasal high flow (nhf) therapy.Air and oxygen are blown from the airvo 2 into the humidification chamber within which water is heated by a heater plate.The gases become humidified in the chamber and then pass into a heated breathing tube, then to the patient interface.Nhf therapy is intended for use with spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases nhf therapy should not be used for life support purposes, and appropriate patient monitoring must be used at all times.Method: the complaint autofill chamber was not returned to fisher & paykel healthcare (f&p) for evaluation as it had been destroyed by the healthcare facility.Our investigation is based on the information and photographs, videography provided by the customer, previous investigations of similar complaints, and our knowledge of the product.Results: visual inspection of the provided photograph and videography revealed a long vertical crack on one of the inlet ports of the autofill chamber, propagating downwards towards the base.Upon device set-up, water was observed to be leaking in large quantities, the issue was identified prior to patient use.Conclusion: without the return of the complaint device, we are unable to confirm the cause of the reported event.Every autofill chamber is pressure tested following the manufacturing process to check for any leaks present in the feed set due to cracks and other causes.Any chamber that fails this inspection is rejected.The subject chamber would have met the required specification at the time of production.Our user instructions that accompany the 900pt561 heated breathing tube and chamber kit state the following: -"do not use the autofill chamber if it has been dropped or been allowed to run dry as this could lead to the chamber over-filling." -"do not use the autofill chamber if the water level rises above the maximum water level line as this may lead to water entering the patient's airway." -"for single patient use only.Reuse may result in transmission of infectious substances.Attempting to reprocess will result in degradation of materials and render the product defective." -"avoid contact with chemicals, cleaning agents, or hand sanitizers.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATED BREATHING TUBE AND CHAMBER KIT
Type of Device
AUTOFILL CHAMBER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
omid taheri
17400 laguna canyon road
suite 300
irvine, CA 92618
8007923912
MDR Report Key17795762
MDR Text Key324019082
Report Number9611451-2023-00884
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K131895
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Distributor
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number900PT561
Device Catalogue Number900PT561
Device Lot Number2102062705
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/24/2023
Initial Date FDA Received09/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/19/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-