VYAIRE MEDICAL, INC 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY
|
Back to Search Results |
|
Model Number 3100A VENTILATOR |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/22/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
H3: 81 other - the customer reported that they will not return the defective unit/part for evaluation.Therefore, no root cause could be determined.However, the customer emailed back and confirmed that he did not find the air connection loose.Additionally, oscillation never stopped and no water was found in any of the colored cap diaphragms.Customer will perform a circuit calibration and performance check, and if everything goes fine, the device can go back to service.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.".
|
|
Event Description
|
It was reported to vyaire medical problem with 3100a ventilator regarding loose air hose connection on back of the unit.It has been indicated that high pitch squealing started coming from the air hose in the back of the oscillator, loose connection, unsafe to adjust while on patient, so equipment was exchanged.Patient tolerated well.No need for manual bagging swift exchange with machines placed next to each other.
|
|