Model Number CLV-190
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported to olympus that the evis exera iii xenon light source power button was broken and right where the light source connects the bottom protector that comes out was broken.The event occurred during reprocessing.There was no patient harm associated with the event.
 
Manufacturer Narrative
To date, the device has not been returned to olympus for evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Corrected field: h4 additional information added to field d8, d9, e2, e3, g2 (added health professional), h3, h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.The device was returned to olympus for inspection, and the customer's allegation was confirmed.In addition, the following non-reportable malfunctions were found during the device evaluation: damaged front panel, faulty scope socket (locking mechanism loose), screws were detached, broken, loose, and damaged, and also a non-olympus lamp was found.Based on the results of the investigation, faulty power switch was confirmed during inspection.Moreover, the front panel near the power switch was found damaged.Thus, it was surmised that it was broken due to an external impact.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17798694
MDR Text Key324036774
Report Number3002808148-2023-10215
Device Sequence Number1
Product Code NWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 11/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/29/2023
Initial Date FDA Received09/22/2023
Supplement Dates Manufacturer Received11/14/2023
Supplement Dates FDA Received11/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/05/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-