• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ROADRUNNER NIMBLE HYDROPHILIC WIRE GUIDE; OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ROADRUNNER NIMBLE HYDROPHILIC WIRE GUIDE; OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Model Number N/A
Device Problems Break (1069); Material Separation (1562); Unraveled Material (1664)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/28/2023
Event Type  malfunction  
Event Description
It was reported that a roadrunner nimble hydrophilic wire guide was found to be broken prior to use.A customer provided photo shows separation of the wire tip from the wire guide.The device did not make patient contact.A new same device from same lot was used to complete the procedure.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.E1- customer (person): (b)(6).H3: device has been returned and preliminary evaluation has been performed, however, our investigation is ongoing and device evaluation summary will be included in our follow up report once our investigation has been completed.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Manufacturer Narrative
Event description: it was reported that a roadrunner nimble hydrophilic wire guide was found to be broken prior to use.A customer provided photo shows separation of the wire tip from the wire guide.The device did not make patient contact.A new same device from same lot was used to complete the procedure.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.Investigation ¿ evaluation a visual inspection of the returned device was conducted.A document-based investigation was also performed including a review of complaint history, device history record (dhr), the instructions for use (ifu), and quality control procedures.One roadrunner nimble hydrophilic wire guide was returned in an opened labelled package.The flexible tip was separated from the rest of the wire guide and the coiling wire was partially unraveled.No other damage noticed.A document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record found no non-conformances related to the reported failure mode.A review of complaint history records shows one other related complaint associated with the complaint device lot.There are 100% inspections in place to ensure the surface of the wire guide is smooth and not damaged.Therefore, it is unlikely this issue affects the entire lot.Based on the available information, cook has concluded that the device was manufactured to specification and there is no evidence suggesting nonconforming product exists either in house or in the field.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: instructions for activating hydrophilic coating: the hydrophilic coating on the wire guide is activated by immersion in sterile water or sterile saline solution.1.Prior to using the wire guide, fill a syringe with sterile water or sterile saline solution and attach it to the flushing port on the wire guide.2.Inject enough solution to wet the wire guide surface entirely.This will activate the hydrophilic coating.The customer reported finding the damage after opening the packaging.A cause for the complaint could not be established.Per the quality engineering risk assessment, no further action is required.The appropriate personnel have been notified, and we will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROADRUNNER NIMBLE HYDROPHILIC WIRE GUIDE
Type of Device
OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key17799655
MDR Text Key324050992
Report Number1820334-2023-01271
Device Sequence Number1
Product Code OCY
UDI-Device Identifier00827002186293
UDI-Public(01)00827002186293(17)260206(10)15210231
Combination Product (y/n)N
PMA/PMN Number
K082536
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberRFSPC-035145-0-I-AQ
Device Lot Number15210231
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2023
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/31/2023
Initial Date FDA Received09/22/2023
Supplement Dates Manufacturer Received11/06/2023
Supplement Dates FDA Received12/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/06/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age53 YR
Patient SexFemale
-
-