Device evaluated by manufacturer: the device was returned for analysis.The main coil was returned inside the introducer and the pusher wire also return for the analysis; it was necessary to make cuts to free the coil.Visual and microscope inspection revealed that the main coil was stretched, detached and bent at the coil arm section and the pusher wire was bent.No more damages were observed.The functional inspection could not be performed, because the main coil and the pusher wire are not interlocking.Dimensional inspection on the main coil revealed that the zap tip od - outer diameter (max) and the primary coil od were within specification.
|
Reportable based on device analysis completed on 04sep2023.It was reported that the coil stuck in the middle of the catheter.The target lesion was located in the spermatic vein.A 4mm x 15cm interlock coil was selected for use.The device was continuously flush before introduction of the coil.During the procedure, a microcatheter was used to release two coils, however, upon deploying the second coil it got stuck in the middle of the catheter and could not be advance despite multiple attempts.When the coil was removed from the catheter, it was noted that the coil got stretched.The procedure was completed with another of same device.No complications were reported, and patient was stable post procedure.However, device analysis revealed that the main coil detached at the coil arm section.
|