Model Number 37603 |
Device Problems
Break (1069); Failure to Deliver Energy (1211); Ambient Noise Problem (2877)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
D10: section d information references the main component of the system.Other relevant device(s) are: product id: 3389-40.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that a patient complained that after an ultrasound examination of the right and left shoulder, their implantable neu rostimulator (ins) turned off and the lead separated.The rep indicated that the procedure of the examination was the same as other types of ultrasound. it was performed using a device that emits ultrasound through a special head and guided over the surface of the patient's body.The reflected wave, returning to the device, creates an image of the examined tissues and organs on the screen, making it possible to analyze the correctness of their construction and determine whether the lesion is inflammatory, degenerative or cancerous.The rep indicated that based on their information, diagnostic ultrasound poses no electrical interference problems for the dbs system, however the transducer should not be placed directly over the ins.No patient symptoms were reported.
|
|
Search Alerts/Recalls
|
|
|