ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY RF DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number PM2240
Device Problems Pacemaker Found in Back-Up Mode (1440); Therapy Delivered to Incorrect Body Area (1508)
Patient Problems Fatigue (1849); Discomfort (2330)
Event Date 09/04/2023
Event Type  Injury  
Event Description
It was reported that the patient presented to the emergency room with symptoms of extra cardiac stimulation and fatigue.The pacemaker (pm) was found to be in backup mode.The physician elected to explant and replace the pm.The patient was discharged following the procedure.
 
Manufacturer Narrative
The reported events of device were in backup mode and extra cardiac simulation were confirmed.The device was received with normal telemetry communication and in backup mode.Visual inspection of the header attachment area detected a bonding anomaly.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.A device hermeticity breach was observed, consistent with feedthrough damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to the internal electronics.The device was cut open to enable further testing and battery was found at elective-replacement level.Hybrid circuitry was tested, indicating high current drain, consistent with moisture damage, depleting the battery prematurely and resulting in the reported events.A manufacturing anomaly may have occurred, which resulted in the header bonding anomaly.The device is included in the assurity and endurity pacemakers header anomaly advisory issued by abbott on 15 march 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY RF DR
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17803867
MDR Text Key324106036
Report Number2017865-2023-46777
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date 11/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2017
Device Model NumberPM2240
Device Lot NumberA000019709
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/04/2023
Initial Date FDA Received09/22/2023
Supplement Dates Manufacturer Received10/27/2023
Supplement Dates FDA Received11/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/17/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1530-2021
Patient Sequence Number1
Treatment
TENDRIL RA LEAD.; TENDRIL RV LEAD.
Patient Outcome(s) Required Intervention;
Patient Age60 YR
Patient SexMale
-
-