• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3013-53
Device Problems Erratic or Intermittent Display (1182); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/15/2023
Event Type  malfunction  
Event Description
It was reported that before use on a patient, the cs100 intra-aortic balloon pump (iabp) unit has a system failure and the display flickering.There was no patient involvement reported.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Manufacturer Narrative
Updated fields - b4, d9, g3, g6, h2, h3, h4, h6 (type of investigation, investigation findings, component codes, investigation conclusions), h10.Additional poc : (b)(4).A getinge field service engineer (fse) observed that system is showing system failure when it¿s starting auto fill procedure.Also observed that display is flickering, problem found with manifold drive assembly and display cable need to change.Replaced new manifold drive assembly and refixed the display cable then checked the system it is working fine.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key17804681
MDR Text Key324129887
Report Number2249723-2023-04186
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3013-53
Device Catalogue Number0998-00-3013-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/15/2023
Initial Date FDA Received09/22/2023
Supplement Dates Manufacturer Received03/26/2024
Supplement Dates FDA Received03/28/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2008
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.
-
-