OLYMPUS WINTER & IBE GMBH INNER SHEATH, FOR 26 FR. OUTER SHEATH; RESECTOSCOPE Back to Search Results
Model Number A22040A
Device Problems Material Deformation (2976); Scratched Material (3020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported to olympus, the inner sheath, had a damaged ceramic tip.There were no reports of patient harm.
 
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The evaluation found scratching on the inside of the tip beak.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information based on the legal manufacturer's final investigation.Correction to the g3 aware date in the initial medwatch.The aware date should have been 04sep2023.The h6 medical device problem code field was also updated to "3020 - scratched material." a review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the damage to the ceramic beak of the sheath was caused by thermal induced impact, wear and tear, improper handling, or mechanical overload like a fall, shock or similar stress.The event can be detected and prevented by handling the device in accordance with the instructions for use which state: ¿4 before use warning, infection control risk: properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.4.1 inspection and testing: inspecting the product.Visually inspect the product.Make sure that it has: no corrosion, no dents, no scratches, ceramic insulation at distal end: visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning: risk of injury, impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product: if the product is damaged or does not function properly, contact an olympus representative or an authorized service center.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNER SHEATH, FOR 26 FR. OUTER SHEATH
Type of Device
RESECTOSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, deutschland 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, deutschland
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17805043
MDR Text Key324179337
Report Number9610773-2023-02683
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K931995
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22040A
Device Lot Number065W-0075
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/28/2023
Initial Date FDA Received09/23/2023
Supplement Dates Manufacturer Received02/09/2024
Supplement Dates FDA Received02/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
A22026A
-
-