C.R. BARD, INC. (BASD) -3006260740 POWERPORT SLIM IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 1716000J |
Device Problems
Separation Failure (2547); Difficult or Delayed Separation (4044)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/24/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the powerport slim implantable port, chronoflex single-lumen, 6f that are cleared in the us.The pro code and 510 k number for the powerport slim implantable port, chronoflex single-lumen, 6f are identified in d2 and g4.H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiry date: 09/2024) h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
|
|
Event Description
|
It was reported that during a port placement procedure, when the sheath introducer was attempted to be peeled away, the valve of the sheath was allegedly unable to be peeled away.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the powerport slim implantable port, chronoflex single-lumen, 6f that are cleared in the us.The pro code and 510 k number for the powerport slim implantable port, chronoflex single-lumen, 6f are identified in d2 and g4.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one fully peeled 6.5fr peel-apart sheath was received for evaluation.Visual and functional evaluations were performed.Bends and twists were noted throughout the fully peeled 6.5fr peel-apart sheath shafts.The valve caps and valves were still attached to the t-handles on both peeled sheath shafts.Therefore, the investigation is confirmed for the reported difficult or delayed separation and identified separation failure issue, the airguard valve did not break properly in the middle.The definitive root cause could not be determined based upon the available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 09/2024), g3, h6 (device).H11: h6 (method, result, conclusion).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that during a port placement procedure, when the sheath introducer was attempted to be peeled away, the valve of the sheath was allegedly unable to be peeled away.There was no reported patient injury.
|
|