• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BRAEMAR MANUFACTURING, LLC C6 MCOT PPM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BRAEMAR MANUFACTURING, LLC C6 MCOT PPM Back to Search Results
Model Number 02-01804
Device Problems Thermal Decomposition of Device (1071); Melted (1385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The device returned with physical damage on the top of the monitor.The damage appeared to be caused by overheat.The patient never reported any technical or physical failures of the monitor during use.
 
Manufacturer Narrative
The device was returned with physical damage near the camera.The patient did not report any technical issues or malfunctions during use.The device was investigated.Engineering evaluation was able to confirm device melt.No internal device was observed during evaluation.It is most probable that the device melt was caused by an external heat source.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C6 MCOT PPM
Type of Device
C6 MCOT PPM
Manufacturer (Section D)
BRAEMAR MANUFACTURING, LLC
1285 corporate center drive
suite 150
eagan MN 55121
Manufacturer (Section G)
BRAEMAR MANUFACTURING, LLC
1285 corporate center drive
suite 150
eagan MN 55121
Manufacturer Contact
beverly okoh
1285 corporate center drive
suite 150
eagan, MN 55121
MDR Report Key17808640
MDR Text Key324151399
Report Number2133409-2023-00059
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081444
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number02-01804
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2023
Initial Date Manufacturer Received 09/07/2023
Initial Date FDA Received09/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/07/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age64 YR
Patient SexMale
-
-