RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Infusion or Flow Problem (2964)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/19/2023 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint on the v60 ventilator indicating that there was an oxygen source failure.The device was reported to be outside of use at the time of the reported problem.No patient harm reported.Investigation is ongoing.
|
|
Manufacturer Narrative
|
E1 reporting institution name, address, city, state, postal code, phone number, reporter phone number: (b)(6).
|
|
Manufacturer Narrative
|
It was reported from the authorized service provider (asp) that the hospital determined that their internal central oxygen supply system pressure was unstable.The hospital re-adjusted their central oxygen supply system, and the equipment began to operate normally.No part replacement was required.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|
|