Model Number LTF-S190-5
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported to olympus, the image of the endoeye flex deflectable videoscope had static.The issue was found during preparation for use, and the procedure was completed using a similar device.There were no reports of patient harm.
 
Manufacturer Narrative
E1 establishment name: (b)(6) hospital.The device was returned and evaluated, and the customer¿s allegation of image was static was not confirmed.Device evaluation found that due to damage on charged coupled device unit, image noise occurred and temporarily the image could not be seen, due to deformation of video connector and case, water tightness was lost, due to damage on insertion pipe, insulation resistance value did not meet the standard value, adhesive on bending rubber section had a chip, due to deformation of bending tube, bending angle in down direction exceeded the standard value, due to damage on the forceps elevator lever, the bending section could not be fixed firmly, due to damage on charged coupled device unit, and a noisy image occurred during angulation in up and down directions.A review of the device history record found no deviations that could have caused or contributed to the reported issue.There was no indication that the event was caused by a misuse or that the event was related to design of the device.Repair history was reviewed and no issues were found related to the reported event.The device was noted to be previously repaired in november 2022.Although a definitive root cause of the defect could not be identified, it was determined that the defect was likely caused due to breakage of the image sensor unit including disconnection by stress of repeated use, external factors or handling, or that the components of the integrated circuit chip and capacitor mounted on the electric circuit board had a defect.If additional information becomes available at a later date, this report will be supplemented.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17815923
MDR Text Key324227324
Report Number9610595-2023-14033
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04953170434648
UDI-Public04953170434648
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K201832
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLTF-S190-5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/31/2023
Initial Date FDA Received09/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/03/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-