Model Number N/A
Device Problem Difficult to Remove (1528)
Patient Problem Pain (1994)
Event Date 07/03/2023
Event Type  malfunction  
Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.G4 - pma/510(k) #: exempt.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
It was reported the stiffening cannula of a ultrathane dawson-mueller mac-loc locking loop multipurpose drainage catheter was difficult to remove.After placing the drainage catheter, the stiffening cannula could not be removed.The cannula was forcefully withdrawn, resulting in the patient experiencing strong pain.The procedure was completed by using another, like device.No other adverse effects nor additional procedures were reported for this incident.
 
Event Description
No additional information regarding patient and/or event details has been received since the previous medwatch report was sent.
 
Manufacturer Narrative
This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Correction: d2a, d2b, g4 d2a: gbo catheter, nephrostomy, general & plastic surgery; lje catheter, nephrostomy d2b: gbo, lje g4: k173035 investigation ¿ evaluation it was reported the stiffening cannula of a ultrathane dawson-mueller mac-loc locking loop multipurpose drainage catheter was difficult to remove.After placing the drainage catheter, the stiffening cannula could not be removed.The cannula was forcefully withdrawn resulting in the patient experiencing strong pain.The procedure was completed by using another, like device.No other adverse effects nor additional procedures were reported for this incident.Reviews of documentation including the complaint history, device history record (dhr), quality control procedures, specifications, manufacturing instructions (mi), and instructions for use (ifu), as well as a visual inspection, functional test, and dimensional verification of the returned device, were conducted during the investigation.One used catheter as well as the supplied disposable stiffening cannula was returned to cook for evaluation.During table top testing, the stiffening cannula was able to be inserted and removed from the catheter, encountering some resistance.Dimensional analysis confirmed that the device and components were manufactured to the correct specifications and tolerances.Upon visual inspection, the device had no visible evidence of surface defects.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhrs for the reported complaint device lot and the related subassembly lots revealed relevant non-conformances which were scrapped prior to further processing of the order.All remaining devices in the lot underwent 100% verification.A database search did not identify any other events associated with the reported device lot.Cook also reviewed product labeling.The product ifu, [t_multi_rev5] ¿multipurpose drainage catheter,¿ provides the following information to the user related to the reported failure mode: precautions: ¿when inserting a stiffening cannula into a catheter with retention suture, hold suture during cannula insertion to avoid bunching or tangling of suture.¿ instructions for use: ¿under fluoroscopic control, perform standard techniques for placement of percutaneous draining catheters, either by seldinger access or trocar access.-once catheter is in desired location, remove any wire guides, trocars, or stiffeners, allowing the catheter to for its configuration.¿ how supplied: ¿supplied sterilize by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened or undamaged.Do not use the product if there is doubt as to whether the product is sterile.Store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred.¿ the information provided upon review of the dmr, ifu and dhr suggests that the device was not manufactured out of specification, and that there are no nonconforming devices in house or out in the field.Based on the information provided, inspection of the returned device, and the results of the investi per the risk assessment no further action is required.This report is required by the fda undergation, cook concluded the cause of this event can be traced to component failure without a manufacturing or design deficiency.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE DAWSON-MUELLER MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER
Type of Device
GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC.
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key17818795
MDR Text Key324278767
Report Number1820334-2023-01289
Device Sequence Number1
Product Code GBO
UDI-Device Identifier00827002097056
UDI-Public(01)00827002097056(17)250302(10)14573004
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Distributor
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/15/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberULT8.5-38-25-P-5S-CLDM-HC
Device Lot Number14573004
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/20/2023
Initial Date FDA Received09/26/2023
Supplement Dates Manufacturer Received11/28/2023
02/01/2024
Supplement Dates FDA Received12/05/2023
02/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/02/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age55 YR
Patient SexMale
-
-