• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. GYNECARE TVT OBTURATOR; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. GYNECARE TVT OBTURATOR; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC Back to Search Results
Catalog Number 810081
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/11/2023
Event Type  malfunction  
Event Description
It was reported that a patient underwent an urological procedure on (b)(6) 2023 and mesh was used.The mesh was broken when it was opened.Another device was used to successfully complete the procedure.There were no adverse patient consequences.No further information was provided.
 
Manufacturer Narrative
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 analysis summary: neuchâtel team received for evaluation a tvto device (product code 810081), batch 3943164.The product was decontaminated and well packaged.The received device was manipulated as original packaging is missing.It was returned some parts of product: part of blue mesh, needles, helical passers, winged gide, lid (from packaging).All returned components have a lot of organic matter around.The mesh was stretched and cut.One needle has been removed from helical passer and it can be observed some damages in the needle tip.No other damages were observed.The observations made during the product evaluation are not aligned with the event description "mesh is broken when it is opened", as the product was manipulated and used.Therefore, the complaint is not confirmed.Events of this type are trended regularly.A review of the batch manufacturing records was conducted, and no related non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYNECARE TVT OBTURATOR
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON SARL-NEUCHATEL
puits-godet 20
neuchatel 2000
SZ   2000
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key17818810
MDR Text Key324278977
Report Number2210968-2023-07063
Device Sequence Number1
Product Code OTN
UDI-Device Identifier10705031000346
UDI-Public10705031000346
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K033568
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2023
Device Catalogue Number810081
Device Lot Number3943164
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2023
Initial Date Manufacturer Received 09/25/2023
Initial Date FDA Received09/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-