Model Number 866389 |
Device Problems
No Audible Alarm (1019); Use of Device Problem (1670)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 09/01/2023 |
Event Type
malfunction
|
Event Description
|
The customer reported that there was a delay in treatment on (b)(6) 2023 at approx.0300 hrs, rm #3119/3 south.Spo2 alarms were allegedly turned off by clinical staff on mp70 ivpm.The customer is requesting onsite support for philips to investigate the incident by pulling logs from ivpm/picix and evaluated by q/r.
|
|
Manufacturer Narrative
|
Philips is in process of obtaining additional information.A final report will be submitted upon completion of the investigation.
|
|
Manufacturer Narrative
|
The following functional tests were performed: spo2 alarm was not going off for patient and patient went into low oxygen alarm and no alarm sound from the monitor was heard.The field service engineer (fse) later found that the customer was turning off spo2 alarms on certain monitors.Fse retrained the customer on devise use.Delay in treatment was reported but no patient info was known or made available at this time.The device was confirmed to be operating per specifications and no failure was identified.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|
|