• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS PATIENT INFORMATION CENTER IX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS PATIENT INFORMATION CENTER IX Back to Search Results
Model Number 866389
Device Problems No Audible Alarm (1019); Use of Device Problem (1670)
Patient Problem Insufficient Information (4580)
Event Date 09/01/2023
Event Type  malfunction  
Event Description
The customer reported that there was a delay in treatment on (b)(6) 2023 at approx.0300 hrs, rm #3119/3 south.Spo2 alarms were allegedly turned off by clinical staff on mp70 ivpm.The customer is requesting onsite support for philips to investigate the incident by pulling logs from ivpm/picix and evaluated by q/r.
 
Manufacturer Narrative
Philips is in process of obtaining additional information.A final report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
The following functional tests were performed: spo2 alarm was not going off for patient and patient went into low oxygen alarm and no alarm sound from the monitor was heard.The field service engineer (fse) later found that the customer was turning off spo2 alarms on certain monitors.Fse retrained the customer on devise use.Delay in treatment was reported but no patient info was known or made available at this time.The device was confirmed to be operating per specifications and no failure was identified.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATIENT INFORMATION CENTER IX
Type of Device
PATIENT INFORMATION CENTER IX
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key17819380
MDR Text Key324286172
Report Number1218950-2023-00693
Device Sequence Number1
Product Code MHX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153702
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866389
Device Catalogue Number866389
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/01/2023
Initial Date FDA Received09/26/2023
Supplement Dates Manufacturer Received09/01/2023
Supplement Dates FDA Received10/20/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-