This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and to correct h4.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, physical stress was applied to the bending section during device handling by the user.Subsequently, the bending tube was damaged.The user can detect the suggested event by handling the device in accordance with the following instructions for use (ifu) section: urf-p6/p6r operation manual, chapter 3 preparation and inspection : 3.3 inspection of the endoscope_ inspection of the bending mechanisms.The user can reduce/prevent the suggested event by handling the device in accordance with the following ifu: urf-p6/p6r operation manual important information ¿ please read before use_ warnings and cautions."do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, or control section.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, or light guide cable with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.".Chapter 4 operation 4.1 precautions."do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.:do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged.".Olympus will continue to monitor field performance for this device.
|