• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. URETERO-RENO FIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. URETERO-RENO FIBERSCOPE Back to Search Results
Model Number URF-P6
Device Problem Material Protrusion/Extrusion (2979)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported to olympus that the uretero-reno fiberscope had an insertion tube leak and fibers broken.The issue occurred during preparation for use.The device was returned for evaluation and during the evaluation the following reportable malfunction was found: there was a metal braid protruded from the distal end rubber due to the broken bending section.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
The device was returned to olympus for evaluation, and the customer¿s allegation was confirmed.The evaluation found the following reportable issues: there was a metal braid protruded from the distal end rubber due to the broken bending section.Other issues found were: broken bending section and air leak from distal end rubber and there were blackspots in the image due to a broken image guide bundle.The investigation is ongoing and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and to correct h4.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, physical stress was applied to the bending section during device handling by the user.Subsequently, the bending tube was damaged.The user can detect the suggested event by handling the device in accordance with the following instructions for use (ifu) section: urf-p6/p6r operation manual, chapter 3 preparation and inspection : 3.3 inspection of the endoscope_ inspection of the bending mechanisms.The user can reduce/prevent the suggested event by handling the device in accordance with the following ifu: urf-p6/p6r operation manual important information ¿ please read before use_ warnings and cautions."do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, or control section.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, or light guide cable with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.".Chapter 4 operation 4.1 precautions."do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.:do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged.".Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17821326
MDR Text Key324330841
Report Number9610595-2023-14098
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170340802
UDI-Public04953170340802
Combination Product (y/n)N
Reporter Country CodeTC
PMA/PMN Number
K172298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/06/2023
Initial Date FDA Received09/26/2023
Supplement Dates Manufacturer Received10/13/2023
Supplement Dates FDA Received10/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/09/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-