The device charging port is physically damaged and needs to be repaired.To correct the issue the front case assembly got replaced.Device was repaired with parts supplied by the bu.Touch screen, nbp, and co2 got calibrated.Device functions are working correctly.The reported problem was confirmed.The cause of the reported problem was not established.A review of the risk management file was performed, and the potential severity is s2 has been identified in the risk document.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.The device was operational after repairs were completed.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|