Model Number 00-1024-R
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/05/2022
Event Type  malfunction  
Manufacturer Narrative
The device charging port is physically damaged and needs to be repaired.To correct the issue the front case assembly got replaced.Device was repaired with parts supplied by the bu.Touch screen, nbp, and co2 got calibrated.Device functions are working correctly.The reported problem was confirmed.The cause of the reported problem was not established.A review of the risk management file was performed, and the potential severity is s2 has been identified in the risk document.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.The device was operational after repairs were completed.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Event Description
This report is based on information provided by philips personnel and has been investigated by the philips complaint handling team.During technical investigation it was noticed that the charging port is damaged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPUS PRO PATIENT MONITOR, WITH PRINTER OPTION
Type of Device
MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)
Manufacturer (Section D)
REMOTE DIAGNOSTIC TECHNOLOGIES LTD.
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 6XW
UK  GU14 6XW
Manufacturer (Section G)
REMOTE DIAGNOSTIC TECHNOLOGIES LTD.
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 6XW
UK   GU14 6XW
Manufacturer Contact
tanya deschmidt
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 -6XW
UK   GU14 6XW
MDR Report Key17821525
MDR Text Key324329888
Report Number3003832357-2023-00664
Device Sequence Number1
Product Code MHX
UDI-Device Identifier05060472441027
UDI-Public05060472441027
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201746
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number00-1024-R
Device Catalogue Number989706000051
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 08/05/2022
Initial Date FDA Received09/26/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-