• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INLIVEN; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INLIVEN; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number V285
Device Problem Inappropriate or Unexpected Reset (2959)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/06/2023
Event Type  Injury  
Event Description
It was reported that this pacemaker reverted to safety mode.It was noted that the patient was pacemaker dependent.The patient was admitted to the hospital and a device replacement procedure was scheduled for a future date.Additional information was received that surgical intervention was undertaken.The device was explanted and replaced.No additional adverse patient effects were reported.The return of the product is expected.At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.
 
Event Description
It was reported that this pacemaker reverted to safety mode.It was noted that the patient was pacemaker dependent.The patient was admitted to the hospital and a device replacement procedure was scheduled for a future date.Additional information was received that surgical intervention was undertaken.The device was explanted and replaced.No additional adverse patient effects were reported.The return of the product is expected.At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.The product has been received for analysis.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode and had undergone resets and that bradycardia therapy remained available.The system resets occurred during a telemetry session and caused the device to enter safety mode.Engineers determined that this device was demonstrating behavior consistent with a high internal cell impedance within the battery.The high internal impedance resulted in the system resets due to temporary high-power consumption and subsequent reversion to safety mode operation.Boston scientific has issued a field safety notice regarding ingenio extended life (el) and cardiac resynchronization therapy pacemaker (crt-p) devices that may exhibit this behavior.This particular device is included in the high impedance in ingenio extended life (el) pacemaker and crt-p advisory population.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INLIVEN
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17822002
MDR Text Key324320108
Report Number2124215-2023-52835
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date 12/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/15/2015
Device Model NumberV285
Device Catalogue NumberV285
Device Lot Number100019
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/26/2023
Supplement Dates Manufacturer Received12/14/2023
Supplement Dates FDA Received12/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/20/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age90 YR
Patient SexFemale
-
-