MEDTRONIC IRELAND ENDURANT II BIFURCATED STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR Back to Search Results
Model Number ETBF2813C145EE
Device Problem Leak/Splash (1354)
Patient Problems Hemorrhage/Bleeding (1888); Hypovolemic Shock (1917); Rupture (2208)
Event Date 09/21/2023
Event Type  Death  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
An endurant ii stent graft and endurant cuff were implanted during the endovascular treatment of a aaa.It was reported during the index procedure the endurant ii stent graft was implanted normally.During the final contrast study a small type ia endoleak was observed.The physician decided to implant an endurant ii aortic cuff.During removal of the delivery system the usual maneuvers were attempted to remove the delivery system however it became blocked in the middle and distal part of the stent graft.The physician rotated the delivery system and managed to remove it however the tapered tip detached and remained inside the patient's body.It was reported the patient expired on an unknown date due to complications post endograft tip rupture.The cause of the type ia endoleak and tip detachment were not provided.
 
Manufacturer Narrative
B5; additional information was received: sizing was noted as -(neck wide 22mm proximal and 28mm distal, neck length 15mm, ifu angle neck, aneurysm wide (7-8cm).Tortuous access and aortic neck.No calcium was noted.It was reported that the physician used normal force when attempting to remove the delivery system.The physician tried to remove the tip using an endovascular approach before deciding to covert to open surgery.Both grafts were explanted and the tip removed while the patient was still alive.The patient suffered complications post the conversion and expired on the day of the index procedure.The physician was unable to provide a cause of the tip detaching.It was reported the entire tapered tip separated.The physician watched the tapered tip detach under fluoroscopy while the delivery system was being removed.The patient experienced an aortic neck rupture.It was said the delivery system got caught at first in the middle of the stent graft at the flow divider and secondly at the distal iliac branch.It got caught in areas of tortuosity.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information received: it was reported that the cause of the aortic rupture that occurred in the aortic neck post implant of the stent grafts is unknown.The cause of death was due to hypovolemic shock ( massive hemorrhage) at the level of the proximal neck of the aneurysm.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information received; it was clarified that the the tapered tip was retrieved without the wire when it was removed via open surgery.The tapered tip was loose through the ascending aorta for a while.The piece separated was entire tapered tip.The anatomy was noted to be tortuous and slightly stenotic.The delivery system became caught due to tortuosity inside the graft first in the middle of the stent graft at the flow divider and secondly at the distal iliac branch.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II BIFURCATED STENT GRAFT
Type of Device
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key17823627
MDR Text Key324322875
Report Number9612164-2023-04372
Device Sequence Number1
Product Code MIH
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
P100021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 10/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberETBF2813C145EE
Device Catalogue NumberETBF2813C145EE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/25/2023
Initial Date FDA Received09/27/2023
Supplement Dates Manufacturer Received10/01/2023
09/28/2023
09/28/2023
Supplement Dates FDA Received10/04/2023
10/05/2023
10/05/2023
Date Device Manufactured01/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age77 YR
Patient SexMale
-
-