|
Model Number 8900 |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/15/2023 |
Event Type
Injury
|
Event Description
|
Reportable based on device analysis completed on (b)(6) 2023.It was reported that the tip of the guidewire was stretched during use.A comet ii pressure guidewire was used for cardiac intervention.During the procedure, the tip of the guidewire was stretched, and the tip could not be shaped.After repeated attempts, it still could not reach the distal end of the blood vessel.Another pressure guidewire was used, and the procedure was completed.No patient complications were reported, and patient was stable post procedure.However, device analysis revealed that the tip was stretched for 7mm in length and detached at an unknown location.
|
|
Manufacturer Narrative
|
Device evaluated by mfr.: the complaint device was received for product analysis.Returned product consisted of an ffr comet pressure wire with the occ cable.The occ cable, tip, and device shaft were visually and microscopically examined for damage or any irregularities.Inspection of the device revealed that the tip was stretched for 7mm in length and detached at an unknown location (due to the stretching of the tip, the detachment location cannot be measured).The ends of the detachment appeared to be smooth with material pushed up at an edge.The detached portion of the tip was not returned for analysis.The occ cable was then connected to the bench top testing equipment and the coefficient values were confirmed to be programmed.No other issues were identified during the product analysis.Product analysis confirmed the reported event, as the tip was stretched and detached.
|
|
Event Description
|
Reportable based on device analysis completed on 01sep2023.It was reported that the tip of the guidewire was stretched during use.A comet ii pressure guidewire was used for cardiac intervention.During the procedure, the tip of the guidewire was stretched, and the tip could not be shaped.After repeated attempts, it still could not reach the distal end of the blood vessel.Another pressure guidewire was used, and the procedure was completed.No patient complications were reported, and patient was stable post procedure.However, device analysis revealed that the tip was stretched for 7mm in length and detached at an unknown location.It was further reported that the tip of the device was stretched and there is no detached portion.Also, no fracture was noted.Device was completely removed from the patient body.
|
|
Manufacturer Narrative
|
Device evaluated by mfr.: the complaint device was received for product analysis.Returned product consisted of an ffr comet pressure wire with the occ cable.The occ cable, tip, and device shaft were visually and microscopically examined for damage or any irregularities.Inspection of the device revealed that the tip was stretched for 7mm in length and detached at an unknown location (due to the stretching of the tip, the detachment location cannot be measured).The ends of the detachment appeared to be smooth with material pushed up at an edge.The detached portion of the tip was not returned for analysis.The occ cable was then connected to the bench top testing equipment and the coefficient values were confirmed to be programmed.No other issues were identified during the product analysis.Product analysis confirmed the reported event, as the tip was stretched and detached.
|
|
Search Alerts/Recalls
|
|
|