MAUDE Adverse Event Report: SYNTHES GMBH STARDRIVE SCREWDRIVER T25 W/ COUPLING FOR 03.019.019/330MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH STARDRIVE SCREWDRIVER T25 W/ COUPLING FOR 03.019.019/330MM Back to Search Results
Catalog Number 03.019.020
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/02/2023
Event Type  malfunction  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: e3: reporter is a synthes employee.H3, h6: part: 03.019.020.Lot: 8318590.Manufacturing site: werk hagendorf.Supplier: na.Release to warehouse date: 12 april 2013.Expiration date: na.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that scrdriver t25 w/coupl f/03.019.019 l330 was found worn from the distal tip.The observed condition was consistent as an end of life indicator for the device.No other issues were identified.The device exhibits damage consistent with repeated use.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed worn of scrdriver t25 w/coupl f/03.019.019 l330 would have contributed to the complained issue.There is no indication that a design or manufacturing issue has caused the complaint condition.It was determined that the reusable instrument was worn from repeated use and servicing.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed.Screwdriver stardrive t25 (current and manufactured.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from (b)(6) reports an event as follows: it was reported that patient underwent open reduction internal fixation (orif) on (b)(6), 2023, for fracture of proximal humerus with stardrive screwdriver and multiloc humeral nail and endcap.The surgery proceeded smoothly, but it became difficult to insert the end cap at the stage of inserting it.The nail was slightly over-inserted, so the surgeon opted for a 2mm endcap, which could not be inserted.The surgeon verified the orientation of the driver from various angles, and attempted to refasten it, but was unable to insert it.The surgeon was concerned that the endcap would come off, so tried to remove it.However, the end cap was stuck between the head subchondral bone and the nail and could not be pulled out.Eventually, the surgeon managed to extract it.The surgeon again tried to place the endcap again.Again, it became difficult to insert, but after several attempts, the surgeon was finally able to insert the end cap firmly.The surgeon believes that there was a possibility that something was interposed between the end cap and the nail, or that the center of the entry hole and the nail, which should have been coaxial, was slightly misaligned when the handle was removed.The surgeon also expressed dissatisfaction with the screwdriver and suggested that if it had an attachment mechanism for the end cap, even if insertion was difficult, extraction would not have been difficult.The surgery was completed successfully with over forty minutes delay.Patient was stable.Upon manufacturer investigation, it was determined that the screwdriver was found worn from the distal tip.This report is for a stardrive screwdriver t25 w/ coupling for 03.019.019/330mm.This is report 3 of 3 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STARDRIVE SCREWDRIVER T25 W/ COUPLING FOR 03.019.019/330MM
Type of Device
SCREWDRIVER
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES HAEGENDORF GMBH-CN
im bifang 6
haegendorf 4614
SZ   4614
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key17825045
MDR Text Key324359809
Report Number8030965-2023-12208
Device Sequence Number1
Product Code HXX
UDI-Device Identifier07611819427204
UDI-Public(01)07611819427204
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.019.020
Device Lot Number8318590
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/12/2023
Initial Date FDA Received09/27/2023
Supplement Dates Manufacturer Received09/28/2023
Supplement Dates FDA Received09/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/12/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNK - END CAPS; UNK - NAILS: MULTILOC HUMERAL
-
-