Model Number 408719
Device Problems High impedance (1291); Ambient Noise Problem (2877); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/17/2023
Event Type  malfunction  
Event Description
Home monitoring detected high lv pacing impedance (>3000 ohms).Some noise-like episodes were also seen in the lv endocardial ecg and lead dislodgement is suspected.Lead currently remains implanted.Should additional information be received, this file will be updated.
 
Manufacturer Narrative
Lead was received for analysis.Explant date is currently unknown.Upon receipt, the lead was found cut 38 cm distal to the is4 connector pin into two segments.The lead was probably cut during the extraction procedure.The analysis showed that the wire of the conductor coil (which leads from the is4 connector to the lead tip) was found fractured 43 cm proximal to the lead tip, which is assumed to be the root cause of the clinical observations.Based on the characteristics, as well as the location of the fracture, it is reasonable to assume that the lead was subject to excessive mechanical forces as the result of clavicular-first rib entrapment.Furthermore, the insulation showed some slight damages, which most likely resulted from the extraction procedure.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENTUS PROMRI OTW QP L-85/49
Type of Device
LV LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key17830805
MDR Text Key324442720
Report Number1028232-2023-04849
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
PMA/PMN Number
P070008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2023
Device Model Number408719
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/27/2023
Initial Date FDA Received09/28/2023
Supplement Dates Manufacturer Received01/19/2024
Supplement Dates FDA Received01/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/06/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-