SIEMENS HEALTHCARE DIAGNOSTICS, INC. ADVIA CENTAUR RUBELLA G; ENZYME LINKED IMMUNOABSORBENT ASSAY, RUBELLA
|
Back to Search Results |
|
Model Number N/A |
Device Problem
False Positive Result (1227)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/05/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A customer from the united states contacted siemens on 2023-09-05 due to observations of imprecision in calibration and quality control (qc) results for advia centaur rubella g (rub g), lot 246.Specifically, the customer noted imprecision for the level-1 control and repeatedly high cvs for calibration.Siemens service undertook troubleshooting with the customer, including a precision study.Service was sent to the customer site to perform checks and decontamination.Following the service intervention, the customer ran qc with acceptable results before and after successfully recalibrating the assay successfully on 2023-09-19.Patient samples were tested the same day, and one sample was observed to have produced an elevated (positive) rub g result which was discordant relative to alternate-method testing.Siemens was informed of the discordant observation on 2023-09-21.Siemens is investigating.
|
|
Event Description
|
The customer reported observation of a positive advia centaur rubella g (rub g) result which was discordant relative to alternate-method testing.The initial elevated result was not reported to the physician.The sample was sent to an external reference lab, and a negative result was returned.This result was accepted as correct.There are no allegations of patient harm, changes in treatment, or delays of diagnosis in association with the observed discordance.
|
|
Search Alerts/Recalls
|
|
|