• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION VALVE SET, EM2400; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION VALVE SET, EM2400; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938724
Device Problem Free or Unrestricted Flow (2945)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/28/2023
Event Type  malfunction  
Event Description
It was reported that a em2400 valve set had parenteral nutrition components mixing between ports 16 and 17.This was observed during compounding on an automated compounding device.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
The lot was manufactured from july 26, 2022 - july 27, 2022.The device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A functional test was performed by installing the sample onto the compounder, and no leak issues were observed.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALVE SET, EM2400
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt 001 mz 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17831175
MDR Text Key324444020
Report Number1416980-2023-04900
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477183
UDI-Public(01)00085412477183
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K002705
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberH938724
Device Lot Number60385258
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/04/2023
Initial Date FDA Received09/28/2023
Supplement Dates Manufacturer Received10/05/2023
Supplement Dates FDA Received10/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-