The device was not received for evaluation.A review of the device history record (dhr) was conducted which confirmed that the device met all quality criteria and manufacturing specifications prior to release.The nxstage system one user guide outlines risks associated with performing hemodialysis therapy and warns all treatments must be administered under a physician's prescription and must be performed by a trained and qualified person, considered to be competent in the use of this device by the prescribing physician.The device meets requirements for electrical and safety standards as outlined in the user guide.The user guide contains a precaution to verify the fluid warmer and power cord show no sign of external damage prior to use.If the unit is damaged, follow the product return procedure described in this manual.It warns to plug the warmer into a properly grounded outlet and describes connection for the power cord from a grounded power outlet to the back of the warmer.Per the user guide, proper electrical hookup in full compliance with all applicable codes and device specifications must be maintained.Details of the electrical connections or during use are unknown.
|
A report was received on 05 sep 2023 from a 43 year old female patient with a medical history including end stage renal disease, who stated the express fluid warmer power cord caught on fire on (b)(6) 2023.Additional information was received on 08 sep 2023 from the home therapy nurse (htn) who stated the patient was not in treatment when the event occurred and there was no adverse impact on the patient or operator.Following the event, the patient continues to treat with the nxstage system.
|