AIZU OLYMPUS CO., LTD. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V3
Device Problem No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The device evaluation found the following: the fogging shown on the screen is caused by moisture ingress, and this was identified as the root cause of the customer phenomenon initially rreported; signs of corrosion were noticed in the control unit that is most likely the result of previous moisture penetration; light guide cover glass and the light guide rod lens, are both corroded by moisture ingress and needed to be replced; and the connector plug unit is cracked.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The customer reported to olympus the uretero-reno videoscope, image problem.The issue occurred during an unknown event.The procedure was unknown.There were no reports of patient or user harm associated with this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause of the observed damaged bending section and metal exposure could not be determined, however, it is likely that the observed partial detachment of one cable support from the bending section structure, due to physical stress during user handling/mishandling, led to a perforation which resulted in an air leak as well as an ingress of moisture into the device.The event may be detected/prevented by following the instructions for use which state: urf-v3/v3r operation manual chapter 3 preparation and inspection urf-v3/v3r operation manual - important information ¿ please read before use_ precautions: do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.- chapter 4 operation 4.1 warnings and cautions: operation do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.Do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17835206
MDR Text Key324491393
Report Number9610595-2023-14229
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170435119
UDI-Public04953170435119
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/05/2023
Initial Date FDA Received09/28/2023
Supplement Dates Manufacturer Received10/23/2023
Supplement Dates FDA Received10/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/08/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse