• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2023
Event Type  malfunction  
Event Description
Philips received a complaint by the customer on the v60 indicating that during preventative maintenance (pm), it was observed that the devices volume does not work.It was reported there was no patient involvement at the time the issue was discovered.The customer evaluated the device with the assistance of the remote service engineer (rse) and confirmed the reported problem.The customer states during pm the volume adjust click is missing when adjustment arrow is tapped, volume adjust is functional and no malfunction reported, speakers are within ohm specs, requests troubleshooting, if possible, for cause of silent volume adjustment on screen.The rse advised the customer that the speaker driver resides on the central processing unit (cpu) printed circuit board assembly (pcba) and may need to be replaced to correct the issue and to perform alarms test in performance verification test (pvt) to ensure volume adjustment is functional.Provided option codes and customer enabled them and the issue was resolved.The device passed required performance verification tests per philips standards and was returned to service.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key17839704
MDR Text Key324566244
Report Number2518422-2023-24879
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 09/14/2023
Initial Date FDA Received09/28/2023
Date Device Manufactured03/31/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-