• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC NONE; ENDOSCOPE PLUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC NONE; ENDOSCOPE PLUS Back to Search Results
Model Number 470057-05
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/13/2023
Event Type  malfunction  
Event Description
It was reported that prior to the start of a da vinci-assisted left inguinal hernia repair surgical procedure, the rotational assembly was rough and grindy on the 30 degree endoscope plus.The procedure was completed with no reported injury.Intuitive surgical inc.(isi) made a follow-up attempt to obtain additional information.However, no further details have been received as of the date of this report.
 
Manufacturer Narrative
An investigation was completed to determine the cause of this reported event.Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The 30 degree endoscope plus instrument was analyzed and found to have a dislodged camera adapter and retaining ring.There was also a friction issue with the camera adapter and the scope shaft.The 30 degree endoscope plus failed the transmittance test.Failure analysis also confirmed discoloration on the housing, mechanical damage on the light guide ferrule, and cable damage on zone a.The complaint regarding the rotational assembly sticking was confirmed by failure analysis, which indicates that the device did contribute to the customer reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NONE
Type of Device
ENDOSCOPE PLUS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key17840321
MDR Text Key324564143
Report Number2955842-2023-18824
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier00886874116555
UDI-Public(01)00886874116555
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191736
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 09/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number470057-05
Device Catalogue Number470057
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/28/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-