• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA APOLLO; ANESTHESIA UNITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA APOLLO; ANESTHESIA UNITS Back to Search Results
Catalog Number 8606500
Device Problems Gas Output Problem (1266); Failure to Deliver (2338); Inaccurate Delivery (2339); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/22/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation has just started; results will be provided in a follow-up report.H3 other text : on-going.
 
Event Description
It was reported that the unit stopped ventilating while on a patient.No injury reported.
 
Manufacturer Narrative
According to the log file analysis the ventilator detected an expiratory gas flow during inspiration.Thus, the device alarmed for reinstall vent.A temporary, faulty electronic peep valve (mv1) due to a delayed closure of this pneumatic valve can be assumed as root cause.A sticking/stucking peep valve is detected during automatic post.During automatic ventilation, pressure control might be disturbed, as consequence airway pressures, tidal and minute volumes can deviate from user settings.Depending on the set alarm limits, alarms regarding inspiratory pressure and/or tidal/minute volume will be generated, as well as technical alarms like reinstall vent.Manual ventilation and monitoring functions are not affected in this case.
 
Event Description
It was reported that the unit stopped ventilating while on a patient.No injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APOLLO
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM   23542
Manufacturer Contact
moislinger allee 53-55
lübeck 23542
4518822868
MDR Report Key17841287
MDR Text Key324551920
Report Number9611500-2023-00343
Device Sequence Number1
Product Code BSZ
UDI-Device Identifier04048675229674
UDI-Public(01)04048675229674(11)190829(17)240206(93)8606500-67
Combination Product (y/n)N
PMA/PMN Number
K042607
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 10/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8606500
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/26/2023
Initial Date FDA Received09/29/2023
Supplement Dates Manufacturer Received09/29/2023
Supplement Dates FDA Received10/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/31/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-