This device is classified as import for export, therefore 510k is not applicable.Model eb-1575k is available in the usa with a 510k number k131028.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the shadow in image.Based on the result, we concluded that it was caused due to the moisture condensation in the ccd module.In addition, our technician confirmed that the lcb (light carrying bundle) broken, the light guide cable buckled, the insertion flexible tube crushed, the remote control buttons cut, the distal body worn out, the angulation down angulation decrease, and the angulation up angulation decrease; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|