Section d information references the main component of the system.Other relevant device(s) are: product id 3387s-40; serial# (b)(6); product type lead.Product id 3708660; serial# (b)(6); implanted: (b)(6) 2014; product type extension.Section d information references the main component of the system.Other relevant device(s) are: product id: 3387s-40, serial/lot #: (b)(6); product id: 3708660, serial/lot #: (b)(6), ubd: 21-jan-2018, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
It was reported that an impedance test was run and showed an orange flag and impedance values greater than 5000.The manufacturer representative (rep) inquired if the patient could have an mri.No other information was provided.Additional information was received by the manufacturer representative (rep), who reported the high impedance value was shown at the right side lead/extension.The cause remains unknown, but possibly due to a fall.No actions/surgery are planned to resolve the impedances.The therapy programming is working around high impedances.
|