Catalog Number 451443V2
Device Problem Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The event date and initial reporter address are currently unknown.A review of the manufacturing documentation associated with lot 18145238 presented no issues during the manufacturing process that can be related to the reported event.Additional information is pending and will be submitted within 30 days upon receipt.
 
Event Description
As reported, there was presence of a hole in the catheter wall in the distal end of a 4f 65cm tempo catheter.There was no reported patient injury.The device is being returned for evaluation.
 
Manufacturer Narrative
After further review of additional information received the following sections have been updated accordingly: b5, d2, d9, g1, g3, g6, h1, h2, h3 and h6.As reported, there was presence of a hole in the catheter wall in the distal end of a 4f 65cm tempo catheter.There was no reported patient injury.The product was not returned for analysis.A product history record (phr) review of lot 18145238 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.Without the return of the device or images for analysis, the reported customer event ¿catheter (body/shaft-puncture/cut¿ could not be confirmed.Shipping/handling factors may have contributed to the reported event.Users are trained to inspect for signs of damage prior to and during use.Any product with damage is not to be used.Information for safety is provided in the products labeling with the intent to make the user aware of the risks.According to the instructions for use (ifu), although not intended as a mitigation of risk, ¿this product is designed and intended for single use.It is not designed to undergo reprocessing and re-sterilization after initial use.Reuse of this product, including after reprocessing and/ or re-sterilization, may cause a loss of structural integrity which could lead to a failure of the device to perform as intended and may lead to a loss of critical labeling/use information all of which present a potential risk to patient safety.Store in a cool, dark, dry place.Do not use if package is open or damaged.¿ based on the information available and the phr review, there is no indication that the event is related to the device design or manufacturing process.Therefore, no preventative or corrective actions will be taken at this time.
 
Event Description
As reported, there was presence of a hole in the catheter wall in the distal end of a 4f 65cm tempo catheter.There was no reported patient injury.Additional information was requested but not provided.The device was not returned for evaluation as previously expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPO
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS US CORP.
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
santiago troncoso 808
juarez, chihuahua 
7863138372
MDR Report Key17843087
MDR Text Key324578996
Report Number9616099-2023-06602
Device Sequence Number1
Product Code DQO
UDI-Device Identifier10705032008075
UDI-Public(01)10705032008075(17)250831(10)18145238
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973401
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number451443V2
Device Lot Number18145238
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/05/2023
Initial Date FDA Received09/29/2023
Supplement Dates Manufacturer Received10/10/2023
Supplement Dates FDA Received10/11/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/10/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient EthnicityHispanic
-
-