Device report from synthes reports an event in canada as follows: it was reported that(b)(6) 2023, when the staff was putting back the instruments into the sets, it was noticed that the numbers etched on the label of the instrument where difficult to see.The issue was noticed post-op, in the instrument room, after decontaminating the instrument.It was further noted that a photo provided by the reporter showed that two devices were involved, and were found to be bent at the tips.This report involves one (1) small hexagonal screwdriver with holding sleeve.This is report 2 of 2 for(b)(4).
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6: photo investigation: the photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned photo.Visual analysis of the photo revealed that both of the small hexagonal scrdriver holding slv appear to have the product information etching worn out.The observed condition was consistent as an end of life indicator for the device.Additionally, the distal tips are lightly bent.No other issues were identified.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The device exhibits damage consistent with repeated use.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the small hexagonal scrdriver holding slv would have contributed to the complained issue.There is no indication that a design or manufacturing issue has caused the complaint condition.It was determined that the reusable instrument was worn from repeated use and servicing.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Physical device investigation: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that small hexagonal scrdriver holding slv was found to have the product information etching worn out.Additionally was found the tip stripped.The allegation can be confirmed.No other defect was found.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A dimensional inspection for the small hexagonal scrdriver holding slv was not performed as it is not applicable to the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the small hexagonal scrdriver holding slv would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): part # 314.02.Synthes lot # 2588947.Supplier lot # 2588947.Release to warehouse date: 24 may 2010.Supplier: synthes gmbh oberdorf.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|