SCHILLER AG TEMPUS LS MANUAL; LOW ENERGY DEFIBRILLATOR Back to Search Results
Model Number 00-3020
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/06/2022
Event Type  malfunction  
Manufacturer Narrative
The device was sent to schiller for investigation.Schiller informed that the device was received without the original battery hence they were not able to check if the failure to recognize the battery is reproducible.Schiller repair department checked the device with their own batteries and there was no problem.The reported problem was not confirmed.A review of the risk management file was performed, and the potential severity is s2 has been identified in the risk document.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.The device was confirmed to be operating per specifications and no failure was identified.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Event Description
This report is based on information provided by philips remote service engineer and schiller investigation team and has been investigated by the philips complaint handling team.Philips received a complaint on the tempus ls indicating that device failed to recognize the battery.The battery was full.Device tested with another battery and failed but the device passed manual self test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPUS LS MANUAL
Type of Device
LOW ENERGY DEFIBRILLATOR
Manufacturer (Section D)
SCHILLER AG
altgasse 68
baar 6341
SZ  6341
Manufacturer (Section G)
SCHILLER AG
altgasse 68
baar 6341
SZ   6341
Manufacturer Contact
tanya deschmidt
altgasse 68
baar 6341
SZ   6341
0417664242
MDR Report Key17850653
MDR Text Key324673155
Report Number3003832357-2023-00679
Device Sequence Number1
Product Code LDD
UDI-Device Identifier07613365002737
UDI-Public07613365002737
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number00-3020
Device Catalogue Number989706001681
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/06/2022
Initial Date FDA Received10/02/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-