Model Number V60
Device Problem Volume Accuracy Problem (1675)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/05/2023
Event Type  malfunction  
Event Description
Philips received a complaint by the biomedical (biomed) engineer on the v60 indicating that the tidal volume was showing all zeros.The device was in use on a patient at the time the reported issue was discovered; however, there was no harm to the patient or user.The biomedical engineer (biomed) informed the remote service engineer (rse) that the tidal volume was showing all zeros and also noticed that the average air was reading 4.0 slpm while troubleshooting.The rse provided the biomed with the latest service manual, recommended that the biomed should replace the flow sensor assembly, and provided the biomed with the replacement part number for repair.The biomed informed the rse that the flow sensor assembly would be replaced.Investigation is ongoing.
 
Manufacturer Narrative
Multiple attempts have been made on 02-oct-2023, 09-oct-2023, and 19-oct-2023 to try to obtain further information about this case, but no response was received from the customer.This file is closed and can be reopened if new information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key17856208
MDR Text Key324790209
Report Number2518422-2023-25203
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/05/2023
Initial Date FDA Received10/02/2023
Supplement Dates Manufacturer Received10/26/2023
Supplement Dates FDA Received10/27/2023
Date Device Manufactured10/22/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-