MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX SETS (M); DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX SETS (M); DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955595
Device Problems Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/15/2023
Event Type  malfunction  
Manufacturer Narrative
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.During visual inspection of the provided photographic samples, the leak was observed from the drain bag.The reported condition was verified.The cause of the condition could not be determined; however, the most probable cause was due to the bag being reused.The event occurred ¿several hours after treatment started¿.All accessories provided within the set, including drain bag, are single use products.This means that once used, the drain bag must not be reused (i.E., emptied and reused during the same therapy) and should be discarded.The prismaflex control unit operator's manual indicates to ¿change fluid bags when the appropriate caution alarm occurs (pbp bag empty, replacement bag empty, dialysate bag empty or effluent bag full)¿.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an external fluid leak was observed originating from a split in the waste fluid bag of a prismaflex m100 set several hours into continuous renal replacement therapy.The set was replaced.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX SETS (M)
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17857652
MDR Text Key324768600
Report Number8010182-2023-00379
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414124397
UDI-Public(01)07332414124397
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 10/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955595
Device Lot Number22J0076
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/02/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
PRISMAFLEX MACHINE
-
-