The device was returned to olympus for evaluation and the customer¿s allegation was not confirmed.The investigation noted that no abnormalities could be identified in any of the areas visible.However, when the scope and the clv-s40pro device, ((b)(6).The related record) were connected to each other, it was noted that no smoke was generated, but that it was a blackout without being output.Further, when changed to equipment because of concern, the images appeared, so we confirmed that the otv-s7pro device was abnormal.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or when additional information becomes available.
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 11 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the reported event could not be identified.However, it¿s likely the cause is related to a faulty power supply unit.Olympus will continue to monitor field performance for this device.
|